Fluke 719P-30G/KIT -0.8 bar to 2 bar Pressure Calibrator

Subtotal (1 unit)*

€4,839.00

(exc. VAT)

€5,951.97

(inc. VAT)

Add to Basket
Select or type quantity
Temporarily out of stock
  • 1 unit(s) shipping from 29 September 2026
  • Plus 2 unit(s) shipping from 13 October 2026
Need more? Click ‘Check delivery dates’ to find extra stock and lead times.
Units
Per unit
1 +€4,839.00

*price indicative

Calibration Options:

Each

€5,007.00

(exc. VAT)

€6,158.61

(inc. VAT)

RS Stock No.:
458-135
Mfr. Part No.:
FLK-719P-30G/KIT
Manufacturer:
Fluke
Find similar products by selecting one or more attributes.
Select all

Brand

Fluke

Product Type

Pressure Calibrator

Battery Type

AA Alkaline

Maximum Pressure Measurement

2bar

Minimum Pressure Measurement

-0.8bar

Accuracy

0.0001 bar

Resolution

0.0001 bar

Pressure Port Type

Gauge

Model Number

719P-30G/KIT

Display Type

Backlight

Current Loop Measurement

Yes

Leak Testing

Yes

Power Source

Battery

Weight

1.2kg

Minimum Operating Temperature

-10°C

Maximum Operating Temperature

50°C

Length

70mm

Width

111.8mm

Standards/Approvals

Traceable Certificate of Calibration

Height

241.3mm

Hazardous Area Certification

No

COO (Country of Origin):
US

Fluke Pressure Calibrator, 2bar Maximum Pressure Measurement, 0.0001 Bar Accuracy - FLK-719P-30G/KIT


This pressure calibrator is a portable handheld instrument designed for field calibration and testing of transmitters, switches and gauges, providing both pressure generation and loop calibration functions. It operates from battery power and presents measurements on a large backlit display, allowing users to source, simulate and measure mA signals while also offering temperature measurement capability via an optional RTD probe. The unit supports negative pressure measurement down to -0.8 bar and delivers traceable calibration documentation.

Features and Benefits:


• 0.0001 bar resolution delivers fine pressure adjustment and repeatable results
• 0.0001 bar accuracy supports reference-class transmitter calibration
• Programmemable pump limits prevent over-pressurisation during setup
• Integrated electric pump enables one-handed pressure generation up to 20 bar
• Current loop measurement allows sourcing and measuring of 0-24 mA signals
• Battery powered using AA alkaline cells for field portability

Applications


• Suitable for calibrating high-accuracy pressure transmitters on-site
• Used for functional testing of pressure switches in maintenance rounds
• Ideal for pressure gauge verification in plant inspection tasks
• Can be used for leak testing procedures with non-corrosive gases
• Used with optional RTD probe for combined temperature verification

What pressure connection type does it use?


The instrument accepts a 1/8 in female NPT pressure connection for field hoses and adapters.

How does temperature affect accuracy outside standard conditions?


Add ±0.002% of full scale per °C for operation outside 15 °C to 35 °C when assessing measurement uncertainty.

What are the power requirements and expected pump cycles?


It requires eight AA alkaline batteries and provides a minimum of 1000 pump cycles to 30 psi, 300 cycles to 150 psi and 100 cycles to 300 psi depending on range.

Which media are permissible for pressure measurement?


Pressure measurement is limited to dry Air and non-corrosive gases when using the internal sensor.

What electrical measurement limits apply when simulating mA?


The mA source supports a maximum 1000 Ω load and the mA simulate function must not exceed 26 V dc.